Description
EPRO MMS6220 Vibration monitoring module MMS 6220
EPRO MMS6220 is a dual‑channel shaft eccentricity monitor module of the Emerson MMS6000 TSI system. It works with PR642X eddy‑current sensors to continuously measure static and dynamic shaft eccentricity and record key characteristic data during machine run‑up and run‑down. It provides configurable alarm relays, analog value outputs and raw dynamic signal outputs for shaft bow evaluation. Built‑in diagnostics detect sensor faults and abnormal measurement signals, and all configuration parameters are password locked. Serial interfaces support local commissioning and data transmission to DCS or diagnostic host. This module is widely used for steam turbine eccentricity protection; original manufacturing has been discontinued, and replacement requires identical model and compatible project configuration.
EPRO MMS6220 Vibration monitoring module MMS 6220
| Parameter | Specification |
|---|---|
| Product Type | Dual‑Channel Shaft Eccentricity Monitoring Module |
| Compatible System | EPRO MMS 6000 Machinery Monitoring System (Emerson) |
| Core Function | Dual independent eccentricity measurement for rotating shafts; captures static and dynamic shaft eccentricity, stores characteristic values during run‑up / run‑down; configurable pre‑alarm and danger alarm relays, sensor diagnostics, analog and dynamic signal output for turbine shaft bow monitoring |
| Signal Type | Eddy‑current sensor input (PR642X series), 0–10 V / 4–20 mA characteristic analog output, 0–20 Vpp dynamic AC output, relay alarm contacts, RS232 local configuration port, RS485 bus port |
| Galvanic Isolation | Galvanic separation for sensor circuit, alarm relay contacts and backplane; compliant with EN 50081‑1, EN 50082‑2 EMC standard |
| Power Supply | 18–31.2 V DC, redundant diode‑isolated power inputs |
| Power Consumption | Max 6 W |
| Diagnosis | Power‑on self‑test, sensor cable break / short‑circuit detection, signal plausibility check, watchdog supervision, password protected parameter setting |
| Indicator | RUN, FAULT, PRE‑ALARM, DANGER and channel status LED indicators on front plate |
| Communication | RS232 for configuration, RS485 for host connection and data upload |
| Hot Swap | Non hot‑swappable; rack power must be de‑energized before replacement |
| Operating Temperature | 0 ℃ ~ +65 ℃ |
| Storage Temperature | -30 ℃ ~ +85 ℃ |
| Humidity | 5% ~ 95%, non‑condensing |
| Protection Class | Module IP00, front panel IP21 |
| Certifications | CE, IEC 61000, IEC 68‑2 |
| Mounting | Eurocard DIN 41494, 6 TE / 3 HE plug‑in module for MMS6000 rack |
| Application | Steam turbine, generator and large rotating equipment; shaft eccentricity and shaft bow protection in power plants, petrochemical and metallurgy |
| Remarks | MMS6220; measuring frequency range 0.017–70 Hz (1.02–4200 RPM); designed to pair with PR642X eddy current probes and preamplifiers; original production discontinued, only available as aftermarket spare parts; firmware and configuration must match the existing MMS6000 rack |

- MMS6220